8. WHAT TO DO IF …?
Malfunction Cause/Remedy
… no weight is displayed
during weighing?
The scale has no power supply.
- Check whether the scale is switched on
- Check whether the rechargeable battery
block is inserted
0.00 does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
is displayed?
The voltage of the rechargeable battery block
drops.
- Charge the rechargeable battery block as
soon as possible
bAtt is displayed?
Rechargeable battery block is empty.
- Charge rechargeable battery block
St0p is displayed?
The maximum load has been exceeded.
- Reduce load
is displayed?
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature
Er:X:11 is displayed?
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight-
more evenly
- Restart scales
Er:X:12 is displayed?
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
Er:X:16 is displayed?
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales