Seca 955 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Échelles Seca 955. Seca 955 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1

seca 955

Page 2

10 • 4.2 Symbols in display5DisplayDisplay element for measured results and for device configuration6Battery compartmentAccommodation for the recharge

Page 3 - CONTENTS

Overview • 11English4.3 Information on rating plateText/Symbol MeaningModel Model numberType Type designationSer. No. Serial numberRefer to user manua

Page 4 - 1. FULL CERTIFICATION

12 • 4.4 Menu structureOther functions are available in the menu of the device. This enables you to configure the device perfectly to suit the conditi

Page 5 - 2. DESCRIPTION OF DEVICE

Before you get started … • 13English5. BEFORE YOU GET STARTED …5.1 Scope of delivery, assembledabcdNo. Components Pcs.aScale 1bPower pack 1cRechargeab

Page 6 - 3. SAFETY INFORMATION

14 • 5.2 Establishing the power supplyPower is supplied to the scale with a rechargeable bat-tery block or a power pack (both included in the scope of

Page 7 - Use of measured

Before you get started … • 15EnglishConnect the powersupply unitThe rechargeable battery compartment is charged with the included power pack.WARNING!P

Page 8

16 • 6. OPERATION6.1 Aligning the scale1. Place the scale on firm and even ground.2. Align the scale.– With the foot screws on the front transport rol

Page 9 - 4. OVERVIEW

Operation • 17English6.2 WeighingCAUTION!Patients can be injured if they fallPeople with limited mobility can fall if they try to sit down on the scal

Page 10 - 4.2 Symbols in display

18 • Start weighing 1. Set the wheel brakes of the rear transport rollers.2. Fold the armrest toward the back in order to make it easier for the patie

Page 11

Operation • 19EnglishTare off additionalweight (TARE)You can use the TARE function to prevent the weighing result from being affected by an additional

Page 12 - 4.4 Menu structure

Instruction manual and guaranteeGB

Page 13 - 5. BEFORE YOU GET STARTED …

20 • The display will flash until a stable weight is meas-ured. The weighing value will then be displayed continuously. The symbol (non-calibratable

Page 14 - Insert the rechargeable

Operation • 21EnglishAutomatic switchoverof weighing rangeThe scales are equipped with two weighing ranges. Weighing range 1 ( ) offers a more precise

Page 15 - Connect the power

22 • 6.3 Additional functions (menu)Additional functions are available in the menu for the scales. This allows you to configure the scales perfectly t

Page 16 - 6. OPERATION

Operation • 23English5. To change your selection or call up another sub-menu, keep pressing one of the arrow keys until the required selection is disp

Page 17 - 6.2 Weighing

24 • NOTE:If you enter the value "0", the function will be switched off. The message "Pt" will no longer be displayed.4. Confirm y

Page 18

Operation • 25EnglishActivate acousticsignals (BEEP)You can select whether an acoustic signal is to be emitted whenever a key is pressed or a stable w

Page 19

26 • Restore factory settings(RESET)You can restore the factory settings for the following functions:1. Select the item "Reset" in the menu.

Page 20 - Determine and evaluate

What to do if …? • 27English8. WHAT TO DO IF …?Malfunction Cause/Remedy… no weight is displayed during weighing?The scale has no power supply.- Check

Page 21 - Automatic switchover

28 • 9. MAINTENANCE/RECALIBRATION9.1 Information about maintenance and recalibrationWe recommend having maintenance performed prior to recalibration o

Page 22

Technical data • 29EnglishThe two figures must match for calibration to be valid. If the sticker and calibration counter do not match, recali-bration

Page 23 - Permanently store

• 3EnglishEnglishCONTENTS1. Full certification . . . . . . . . . . . . . . 42. Description of device . . . . . . . . . . 52.1 Congratulations! . .

Page 24 - Activate Autohold

30 • 11.SPARE PARTS12.DISPOSAL12.1 Disposal of deviceDo not discard with household waste. The device must be disposed of properly as electronic waste.

Page 25 - Activate acoustic

Warranty • 31English12.2 Batteries (including rechargeables)Spent (rechargeable) batteries should not be discarded with household waste, regardless of

Page 26 - 7. CLEANING

KonformitätserklärungDeclaration of conformityCertificat de conformitéDichiarazione di conformità Declaratión de conformidad OverensstemmelsesattestF

Page 27 - 8. WHAT TO DO IF …?

955Die nichtselbsttätige PersonenwaageThe non-automatic personal scalesLe pèse-personnes non automatiqueLa bilancia pesapersone non automaticaLa báscu

Page 28 - 9. MAINTENANCE/RECALIBRATION

… entspricht dem in der Bescheinigung über dieBauartzulassung beschriebenen Baumuster. Die Waage erfüllt die geltenden Anforderungenfolgender Richtlin

Page 29 - 10.TECHNICAL DATA

…odpovídá typu popsanému v certifikátu schválenítypu.Váha splňuje platné požadavky těchto směrnic: 2009/23/ES o vahách s neautomatickou činností, 93/4

Page 30 - 12.DISPOSAL

4 • 1. FULL CERTIFICATIONWith products from seca you are not only purchasing technology developed over a century, but also quality that has been valid

Page 31 - 13.WARRANTY

Description of device • 5English2. DESCRIPTION OF DEVICE2.1 Congratulations!With the seca 955 electronic scale, you have just pur-chased a highly prec

Page 32

6 • 3. SAFETY INFORMATION3.1 Basic safety precautionsHandling the instrument • Please take note of the information in this user manual.• Keep the user

Page 33

Safety information • 7EnglishUse of measuredresults• Before you save and re-use values measured with the seca 955 (e.g. in the seca analytics 105 PC s

Page 34

8 • 3.3 Handling (rechargeable) batteriesThis instrument is delivered with a rechargeable battery block. Heed the following safety instructions.WARNIN

Page 35

Overview • 9English4. OVERVIEW4.1 Controls/Features56124378910111213No. Control/Feature Function1To switch scales off and on2Arrow key• While weighing

Commentaires sur ces manuels

Pas de commentaire